Philips Medical Systems Recalls Azurion 7 B20 Systems Due to Unexpected Table Movement

FDA Recalls medium FDA · · United States

Philips Medical Systems Nederland B.V. has issued a voluntary recall for Azurion 7 B20 systems because the table may move unexpectedly during geometry resets, posing a potential safety risk.

What Happened

Philips Medical Systems Nederland B.V. has initiated a voluntary recall of the Azurion 7 B20 System. The recall was issued because, under certain conditions, the system’s table may move unexpectedly when the “Reset Geometry” button is pressed. This unintended movement can occur even when a table lock is currently active.

Which Products Are Affected

The recall affects a total of 1,503 units worldwide, with 326 units distributed across the United States. The following products are included in this recall:

  • Product Name: Azurion 7 B20 System
  • Model Numbers: 722068, 722226, 722236
  • Software Versions: All versions (R1.X, R2.X, R3.X)
  • UDI-DIs: 884838085367, 884838099272, 884838116801
  • Serial Numbers: All serial numbers are affected.

What You Should Do

Healthcare facilities and operators using the Azurion 7 B20 System should exercise caution when using the “Reset Geometry” function. Users should be aware that the table may move despite the activation of a table lock. For further information regarding remediation or software updates, contact Philips Medical Systems Nederland B.V. directly at their headquarters in Veenpluis 4-6, Best, Netherlands.

Why This Matters

Unexpected movement of medical equipment during clinical procedures poses a risk of injury to both patients and medical staff. The failure of the table lock to prevent movement during a geometry reset is a significant safety concern for surgical and diagnostic environments.

Source

Information provided by the FDA under Recall Number Z-1347-2026.

Source: FDA Official Notice

Frequently Asked Questions

What is this fda recalls alert about?
Philips Medical Systems Nederland B.V. has issued a voluntary recall for Azurion 7 B20 systems because the table may move unexpectedly during geometry resets, posing a potential safety risk.
Which agency issued this alert?
This alert was issued by FDA. The original notice is available at the source link at the bottom of this article.
How severe is this alert?
This alert is classified as "medium" severity. Stay informed and follow agency guidance.
What area is affected?
This alert affects United States. Check with FDA for the most current geographic scope.
Where can I find more fda recalls alerts?
Browse all fda recalls alerts on Areazine at areazine.com/recalls/fda/ for the latest updates from FDA and other agencies.